Sep 29, 2025

Public workspaceUroMOCA and StimPod Daily Testing in Pig

  • Dennis Bourbeau1,2,
  • Brett Hanzlicek3
  • 1MetroHealth Medical Center, Cleveland, Ohio;
  • 2Louis Stokes VA Hospital, Cleveland, Ohio;
  • 3Advanced Platform Technology Center, Louis Stokes VA Hospital, Cleveland, Ohio
Icon indicating open access to content
QR code linking to this content
Protocol CitationDennis Bourbeau, Brett Hanzlicek 2025. UroMOCA and StimPod Daily Testing in Pig. protocols.io https://dx.doi.org/10.17504/protocols.io.dm6gp3d2pvzp/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: September 27, 2023
Last Modified: September 29, 2025
Protocol Integer ID: 88485
Keywords: stimpod daily testing in pig, daily testing of stimpod, stimpod daily testing, data from the stimpod, stimpod, pig, uromoca, daily testing, protocol, session
Abstract
This protocol is for the daily testing of StimPods implanted in a pig. Data from the StimPods and UroMOCA are recorded during these sessions.
Troubleshooting
Tolerance Testing
Attach BluMOCA receiver/antenna to harness/jacket
Charge StimPods, then confirm wireless connection to computer
Wake up UroMOCA
Stimulation permutations

20 Hz, 500 Hz, 10 kHz
Stim0, Stim1
Stimulation duration: 1-5 seconds ON, >30 s rest
For each stimulation permutation, determine upper limit of stimulus intensity/amplitude without distress or discomfort to animal
Document subject’s responses
Recruitment Curve
Attach BluMOCA receiver/antenna to harness/jacket
Charge StimPods, then confirm wireless connection to computer
Wake up UroMOCA
Administer stimulation while recording bladder pressure and EMG
Stimulus duration: 5 seconds on, > 30 s rest
Stimulus amplitude range: [0:25 µA:tolerance limit]
Stimulus amplitudes randomized within range
For each stimulation permutation
Within tolerance range for that permutation
Document subject’s responses
Simultaneous Stimulation
Attach BluMOCA receiver/antenna to harness/jacket
Charge StimPods, then confirm wireless connection to computer
Wake up UroMOCA
Stimulus duration: 5-10 seconds on
Amplitude for 20 Hz stimulation: 90% of full recruitment
Amplitude for 500 Hz or 10 kHz stimulation: minimum amplitude (threshold) to achieve maximum EMG reduction
Measure bladder pressure and EMG
Document subject’s responses