| Step | Problem | Possible reason | Solution |
| Procedure 1
Step 12 | Collagen solution does not gel after 60 minutes incubation.
Air pockets form after 60 minutes incubation. | Improper pH of the collagen solution.
The chamber is not sealed tightly.
Insufficient mixing before injection. | Adjust the pH of the mixture.
Change the O-ring seal the chamber tightly.
Keep injecting while the collagen solution leaks into the space between two adjacent vessels.
Briefly pipette to mix thoroughly before injection. |
| Procedure 2
Step 14 | Micro bubbles observed in vessel wall after dehydration. | Aggresive pipetting during mixing. | Mix all the components gently in an Eppendorf tube.
Briefly centrifuge before injecting collagen solution into the molds. |
| Procedure 2
Step 17 | Perfusion chamber cannot be tightened evenly. | The PDMS clamps are too thick. | Trim the PDMS clamps.
Change O-ring. |
| Procedure 2
Step 20 | Additional vessels begin to leak following the repair process. | The extraction of media causes a meniscus to form, resulting in increased vessel leakage. | Disassemble the chamber while keeping the media in place and use absorbent Kimwipes to contain the liquid. |
| Procedure 3
Step 28 | Bubbles inside lumen after rotation. | Medium leaks out or evaporates during rotation. | Fix the chamber orientation so that the mandrels are parallel to the rotator platform.
Fabricate PDMS caps to seal the mandrels after the injection of the endothelial cell suspension. |
| Procedure 4
Step 33.6 | Bubbles keep growing in the TEBV perfusion chamber during perfusion. | The chamber does not seal tightly.
Bubbles in the side loop tubing.
Possible leakage in vessels. | Seal the chamber tightly and evenly.
Remove bubbles as much as possible in the chamber and side loop.
Perform leakage test and replace leaky vessels. |