Create calibration curves for each assay in the correct diluent. Each diluent has been optimized to work for spike and recovery and dilution linearity in both cerebrospinal fluid and plasma. While the calibration curve can be adjusted to any needed concentration range, the following calibration curves work well for biological samples:
The curves below are written for the following stock concentrations of recombinant protein:
CD9 (Abcam ab152262)= 40ug/mL
CD63 (Origene TP301735)= 50ug/mL
CD81 (Origene TP317508)= 120ug/mL
The diluent used for each assay are:
CD9= 30% Quanterix Neuroplex 3A sample diluent (Quanterix 102002) and 70% Sample/Detector Diluent (Quanterix 101359)
CD63= Quanterix Sample Diluent E (Quanterix 101579)
CD81= Quanterix Sample Diluent E (Quanterix 101579)
CD9 in 30% Neuroplex 3A Diluent/70% Sample/Detector Diluent
9) 1.215μL*40μg/mL = 600μL*81ng/mL
8) 200μL*81ng/mL = 600μL*27ng/mL
7) 200μL*27ng/mL = 600μL*9ng/mL
6) 200μL*9ng/mL = 600μL*3ng/mL
5) 200μL*3ng/mL = 600μL*1ng/mL
4) 200μL*1ng/mL = 600μL*333.3pg/mL
3) 200μL*333.3pg/mL = 600μL*111.1pg/mL
2) 200μL*111.1pg/mL = 600μL*37pg/mL
1) 200μL*37pg/mL = 600μL*12.34pg/mL
10) 2μL*50μg/mL = 100μL*1μg/mL
9) 5.4μL*1μg/mL = 600μL*9ng/mL
8) 200μL*9ng/mL = 600μL*3ng/mL
7) 200μL*3ng/mL = 600μL*1ng/mL
6) 200μL*1ng/mxL = 600μL*333.3pg/mL
5) 200μL*333.3pg/mL = 600μL*111.1pg/mL
4) 200μL*111.1pg/mL = 600μL*37pg/mL
3) 200μL*37pg/mL = 600μL*12.3pg/mL
2) 200μL*12.3pg/mL = 600μL*4.11pg/mL
1) 200μL*12.3pg/mL = 600μL*1.37pg/mL
10) 2μL*120μg/mL = 240μL*1μg/mL
9) 5.4μL*1μg/mL = 600μL*9ng/mL
8) 200μL*9ng/mL = 600μL*3ng/mL
7) 200μL*3ng/mL = 600μL*1ng/mL
6) 200μL*1ng/mxL = 600μL*333.3pg/mL
5) 200μL*333.3pg/mL = 600μL*111.1pg/mL
4) 200μL*111.1pg/mL = 600μL*37pg/mL
3) 200μL*37pg/mL = 600μL*12.3pg/mL
2) 200μL*12.3pg/mL = 600μL*4.11pg/mL
1) 200μL*12.3pg/mL = 600μL*1.37pg/mL
Plate each of these putting 360uL in each well of the 96 well plate.
Dilute each sample in the diluent for the assay for which it is being used. Each sample should be diluted at minimum 4x in the respective sample diluent. This dilution can be done by hand or on the HD-1 or HD-X insturnment.