License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
DISCLAIMER – FOR INFORMATIONAL PURPOSES ONLY; USE AT YOUR OWN RISK
The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment. Any action you take or refrain from taking using or relying upon the information presented here is strictly at your own risk. You agree that neither the Company nor any of the authors, contributors, administrators, or anyone else associated with protocols.io, can be held responsible for your use of the information contained in or linked to this protocol or any of our Sites/Apps and Services.
Abstract
This protocol was developed as a semi-high throughput technique to perform phage display immunopreceipitation. Here we describe all the steps: from isolating antibodies from a complex matrix of patient sample, washing the sample and amplify targets through two rounds of selection.
Round 96 well 2ml deep well plate- Genesee Science Cat No: 22-485
Square 96 well 2ml deep well plate- VWR Cat No: 75870-796
12 span 1000uL multichannel pipet
Reagents Needed:
Protein A/G Beads
1xTBS-Tween
BSA Fraction V
LB-Carb
BLT5403
Phage Display Library
RIPA
-10mM Tris-HCL (pH 7.4)
-1.0% Triton X
-0.1% SDS
-140mM NaCl
TNP40
-10mM Tris-HCl (pH 7.4)
-140mM NaCl
-0.1% NP40 (Tergitol)
SM Buffer
-50mM Tris-HCl (pH 7.4)
-100mM NaCl
-10mM MgSO4+7h2O anhydrous (powder)
Protocol materials
Sodium chlorideP212121
Before start
Before getting started you will need a titered phage display library, patient samples and all required materials mentioned in this protocol.
Day 1- Blocking Plates
Day 1- Blocking Plates
Make blocking solution to prevent sample and library from binding to plates. This step also helps with movement of protein A/G beads later on.
Blocking solution (make fresh for every experiment) :
3.0% BSA Fraction V
1x TBS-Tween
Mix thoroughly and filter through a 0.22um filter
Aliquot blocking solution into Round-bottom 96 well 2ml deep well plates according to the number of wells and plates needed for experiment for both rounds of IPs. Seal tightly with foil seal.
Example: If running 4 plates worth of samples (384 samples including controls), block a total of 8 plates, 4 for each round.
Incubate plates at 4 °C inverting overhead or rocking on platform Overnight or for a minimum of 04:00:00 at Room temperature. Solution can block for several days at 4 °C.
4h
Day 1- Setting up IPs
Day 1- Setting up IPs
After plates have sufficiently blocked, unseal plates and discard all blocking solution contained in one set of IP plates by dumping the contents into the sink and gently tapping on clean paper towels. Continue blocking the other set of plates for round two of IP.
Add 500 µL of phage library per well. If a canary well is desired (to control for phage contamination of reagents during the protocol) add SM buffer to the canary well instead of the phage library.
Add sample to library. Amounts will vary depending on sample type and concentration of IgG in sample:
1 µL of serum OR plasma in 1x Storage Buffer
10-20 µL of CSF in 1x Storage Buffer
1 µg of Commercial Antibody (for positive control, recommend to put one of these on each plate randomized location.
An optional but recommended sample type is also an AG control well with no sample to control for nonspecific binding of phage library to A/G beads. We Recommend putting 1-2 of these on each plate in a random location.
Seal plate tightly (should be able to see outline of the individual wells) with foil seal and plastic plate sealer tool. Then, incubate by inverting overhead or on rocker platform at 4 °COvernight (minimum of 6 hours, maximum of 18 hours)
On a clean, phage-free lab bench, prepare an overnight starter culture of E.coli (BLT5403) as a working stock from a frozen glycerol stock in LB-Carb.
Day 2- Preparing AG Beads
Day 2- Preparing AG Beads
Prepare A/G Beads:
NOTE: handle the beads gently using wide bore tips if possible to prevent damaging the beads.
Fully re-suspend beads by gently flipping by hand or placing on overhead rotator.
Calculate amount of beads needed. For the peptidome library we use 10 µL of A and 10 µL of G beads per reaction however some prefer to use 20 µL of A and 20 µL of G per reaction to increase the capture of IgG and IgA.
Mix packed Protein A and Protein G beads in a 1:1 ratio.
Aliquot amount of beads needed into a container able to hold more twice the volume of beads. We typically use a 5mL eppendorf tube or 15mL conical tube. Add 1x bead volume of cold TNP40 and mix overhead for 00:05:00 at 4 °C or until fully resuspended. Place on magnetic rack until beads fully pelleted.
5m
Remove supernatant and immediately add cold TNP40 again. Mix overhead again. Repeat this wash step a total of three times to fully remove the solution that the beads were stored in.
On the final wash step, elute the beads in enough TNP40 for the number of reactions needed. From experience, eluting in 25 µL per reaction works well to form a nice droplet with the widebore tips when adding beads to the peptidome and sample.
Day 2- Adding Beads to Sample Bound to Library Targets
Day 2- Adding Beads to Sample Bound to Library Targets
2m
2m
Spin plates from overnight incubation at 800 rpm for 00:02:00 to remove liquid off the seal.
Gently remove seal.
2m
Put washed A/G beads into a reservoir container.
Add A/G beads to sample/library plate with wide bore tips using a multichannel quickly, making sure the beads stay resuspended and do not settle on the bottom of the reservoir.
Seal plate with foil seal.
Overhead mix for 01:00:00 at4 °C
NOTE: Don't exceed 1 hour, more time will increase nonspecific binding. (50min-1hr of bead binding is okay; start first group of plates at 50 mins)
1h
Simultaneously, prepare new E.coli culture by growing 1-3mL of the overnight culture in enough LB-Carb (0.5mL for each sample).
Day 2- Immunoprecipitation Washes
Day 2- Immunoprecipitation Washes
1m
1m
NOTES:
Keep RIPA on ice to keep samples cool during washes
Plate spins should be at 800 rpm for 00:01:00 on a tabletop centrifuge, as we have found that this is the minimum amount of time and speed required to remove liquid from the top of seal. This may vary by centrifuge.
After hour is up, spin 2 plates (first set if doing more than 2 plates total) at 800 rpm00:01:00
2m
Place plate magnet on top of a firm ice bucket and plate on top of magnet. Wait 1-2 minutes for all of the beads to collect on magnet and liquid to clarify. Remove foil seal to verify that supernatant is clear.
Using a 12-channel P1000 multichannel, aspirate the supernatant from each, discarding the tips.
Using a second multichannel, add 500 µL of RIPA from a reservoir (empty tip box works great) to that row. Don't let the beads stay dry for too long.
Repeat with each rows.
Seal plate very well with foil seal and sealer
Take plate off magnet and invert a few times till beads are resuspended and place on rocker at 4 °Cfor 00:07:00 to 00:10:00 on the rocker so liquid is moving freely in each well.
If doing more than one set of plates, begin on the next set of two plates while the first pair incubates on the rocker.
17m
After 00:07:00 to 00:10:00 of incubation, spin plates at 800 rpm for 00:00:20, and place on magnet to let beads collect while carefully removing the seal
Aspirate 500 µL of RIPA from one row of the plate and add 500 µL of RIPA to same row. Repeat for all rows.
Seal plate very well with foil seal and sealer
Take plate off magnet and invert a few times till beads are resuspended and place on rocker at 4 °Cfor 00:07:00 to 00:10:00.
34m 20s
Repeat step18 three more times for a total of 5 washes.
NOTES:
Each addition of RIPA counts as a wash.
Aim for no more than 2 hours for the 5 washes such that the sample is not in RIPA for too long.
When doing multiple plates, batch them in pairs for easy centrifugation. i.e. six plates would be done in three pairs at a time.
Day 2- Lysis/Amplification
Day 2- Lysis/Amplification
After final/5th wash, remove supernatant and add 150uL LB-Carb to each well.
Using a multichannel, mix one row until all beads resuspended and transfer to a new Square 96 well 2ml deep well plate. Discard tips.
Repeat with fresh tips for each row.
Add 500 µL E.coli at OD600 0.3-0.6 to the bead/LB mixture.
NOTES: If doing multiple plates, try to add the E. coli to the LB around the same time so lysis occurs simultaneously.
Seal with a gas permeable seal, and incubate in a shaker incubator for 1-2 hrs.
NOTES:
Catch the clarification as soon as possible!
Compare lysis to a canary well (can be on a completely separate plate or tube) which should remain cloudy while the wells containing phage library clarify when on magnet to pellet beads.
Once the wells look clear, proceed immediately to next step. Do not let it incubate for longer than necessary, as this will decrease phage titers.
Add 5.0 Molarity (M)Sodium chlorideP212121 to each well for a final molarity of 0.5 Molarity (M) in the clarified E. coli culture (volume changes with amount of lysate, normally around 60-65uL).
Day 3- Preparing for Round 2
Day 3- Preparing for Round 2
30m
30m
Spin plate at 4 °Cfor 00:30:00 at 3220 rcf (or max speed) to remove E. coli cell debris.
While plate is spinning, empty out blocking solution from second round of plates.
After spinning, use the Integra or a multichannel to transfer 500 µL from the top of the wells, avoiding the cell debris at the bottom, to the previously blocked plates.
NOTE: Save Round 1 lysates and store at 4 °C as a backup in case round 2 goes poorly or if you wish to sequence Round 1.
30m
Add samples in same volume and configuration as done in starting Round 1 to plates.
Repeat IP one more time for a total of 2 rounds of immunoprecipitation. One can do additional rounds of IP for more selection if desired.
Store final lysate short term at 4 °C or long term at -20 °C
To prepare samples for sequencing, follow with protocol below: