Public workspaceProtocol for Using the Isolera-Biotage for "Wide-Polarity-Range" (WPR)

  • Ricardo Borges1,
  • Rômulo Pereira de Jesus1
  • 1Federal University of Rio de Janeiro, Brazil.
  • LAABio-IPPN-UFRJ
Icon indicating open access to content
QR code linking to this content
Protocol CitationRicardo Borges, Rômulo Pereira de Jesus 2025. Protocol for Using the Isolera-Biotage for "Wide-Polarity-Range" (WPR). protocols.io https://dx.doi.org/10.17504/protocols.io.eq2ly4jjelx9/v1
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: August 16, 2025
Last Modified: August 19, 2025
Protocol Integer ID: 224795
Keywords: Isolera-Biotage system, Fractionation protocol, Flash chromatography, Complex mixture separation, Polarity-based fractionation, biotage system this method, fractionating complex mixture, method on the isolera, biotage system, complex mixture, biotage, compound, secondary metabolite, applicable to silica gel, isolera, such as secondary metabolite, c18 stationary phase column
Abstract
This protocol describes the procedure for fractionating complex mixtures using the Wide-Range-Polarity (WRP) method on the Isolera-Biotage system. This method employs a controlled elution gradient to separate compounds with a wide range of polarity, such as secondary metabolites from natural extracts. The method is applicable to silica gel or C18 stationary phase columns, depending on the sample characteristics.
Troubleshooting
Materials and Reagents Required
Adsorbent Resin:
  • Silica gel (10 g or 12 g columns)
  • C18-functionalized silica gel (10 g or 12 g columns)
  • HP20-SS (10 g or 12 g columns)

  • Flash chromatography column/cartridge
  • Column frits
  • Sample injection syringe (luer-lock)
  • Glass pipette

  • 16 mL glass tubes with screw caps, properly labeled: 40 units per sample
  • 1.5 mL pre-weighed vials with caps (for mass balance): 40 units per sample
  • Evaporation equipment (Speed-vac) or nitrogen (N₂) flow
  • Rack for organizing tubes and vials
Solvents:
For silica gel columns (normal phase):
  • Solvent 1: Hexane (position 1S)
  • Solvent 2: Ethyl acetate (position 2S)
  • Solvent 3: Methanol (position 4S)

For C18 columns (reverse phase):
  • Solvent 1: Water (position 3S)
  • Solvent 2: Methanol (position 4S)
  • Solvent 3: Ethyl acetate (position 2S)
Equipment:
  • Isolera-Biotage
  • SpeedVac or nitrogen (N₂) flow
  • Solvent-resistant labels
Initial Preparation
Column Preparation:

  • Pack the column.
  • Place the column in the Isolera-Biotage system.
  • Equilibrate the column by passing 3 column volumes (CV) as part of the method.
Sample Preparation:

  • Dissolve the sample (up to 500 mg for a 10–12 g column) in the smallest possible volume of the initial gradient solvent.
  • Ensure the sample is homogeneous and free of particles.
Solvent Setup:

  • Fill the corresponding bottles with the appropriate solvents and connect them to the correct positions on the Isolera-Biotage.
  • Adjust the volumes in each bottle as needed.
Tube Organization:
  • Label the fraction tubes and arrange them in the equipment racks according to the required number.
System Setup
Initialization:
  • Turn on the Isolera-Biotage.
  • Select the “Chemistry” option.
Method Configuration:
  • In “Setup,” enter the solvent volumes in the bottles.
  • Select “Method” → “Open” → “System Owner” (“LAABio”) and choose the corresponding method:
WRP-HEX-ACET-MeOH for silica gel
WRP-C18 for functionalized C18

  • Name the sample and save the method.



Gradient Adjustments:

  • Verify the elution gradient and fractionation conditions.

For silica gel: 100% Hexane → 100% Ethyl acetate → 100% Methanol
For C18: 100% Water → 100% Methanol
Fractionation Execution
Equilibration:

  • Select “Equilibrate.”
  • Wait for the system to pass the required volume of the initial solvent through the column.
  • Complete all initial steps.
Sample Loading:

  • Remove the tubing from the top of the column.
  • Connect the syringe (without the plunger) and load the dissolved sample from the top.
  • Inject the sample directly into the column by selecting “LOAD SAMPLE” to allow flow.
  • After completing the injection, select “CLOSE” to prevent air from entering.
  • Reattach the tubing to the column.



Start of Run:
  • Select "Gradient" to begin the fractionation process.
Completion:
  • At the end, select "TURN LAMP OFF."
  • Remove the tubes containing the fractions and any residual solvents.
  • Remove the column.
  • Copy the chromatogram data to a USB drive.
Completion:
  • At the end, select “TURN LAMP OFF.”
  • Remove the tubes containing the fractions and any residual solvents.
  • Remove the column.
  • Copy the chromatogram data to a USB drive.



Post-Fractionation
Fraction Concentration:
  • Concentrate the fractions using a SpeedVac or N₂ flow until dry, if desired.
Storage:
  • Store the dry fractions in a freezer (-20 °C or -80 °C, if available).
Recording and Updating:
  • Enter the fraction information into the laboratory sample management system.