Perform LC-MS analysis with an Orbitrap Eclipse coupled with a Proxeon EASY-nLC1200 liquid chromatography pump. Separate peptides on a microcapillary column (100 m inner diameter) packed with ~ 35 cm of Accucore150 resin (2.6 m, 150 Å, Thermo Fisher Scientific) with 8-23% (3-73 min), 23-30% (73-80 min), 30-100% (83-86 min) gradient of ACN (0.1% of FA) at 550 nL/min flow rate. For analysis, use multi-notch MS3-based TMT method coupled with Real Time Search algorithm (REF). The scan sequence starts with MS1 spectra analyzed by Orbitrap (resolution 120,000 at 200 Th, 400-1500 m/z, automatic gain control (AGC) target 2105, maximum injection time 50 ms). Monoisotopic peaks should be assigned, use precursor fit filter (70% for a fit window of 0.5 Th), and apply dynamic exclusion window (90 s, ± 10 ppm). Analyze MS2 spectra by quadrupole-ion trap with collision-induced dissociation (Rapid scan rate, AGC 1.0104, isolation window 0.5 Th, normalized collision energy (NBE) 34, maximum injection time 80 ms). Synchronous precursor selection (SPS) API-MS3 scan should collect top 10 most intense b- or y-ions matched with the real-time search algorithm (REF). Fragment MS3 precursors by high energy collision-induced dissociation (HCD) and analyze with the Orbitrap (NCE 45, AGC 2.5105, maximum injection time 200 ms, resolution 50,000 at 200 Th). Set closeout at two peptides per protein for each fraction.