Jan 13, 2023

PLATERO: A Calibration Protocol for Plate Reader Green Fluorescence Measurements

PLATERO: A Calibration Protocol for Plate Reader Green Fluorescence Measurements
  • 1Universitat Politecnica de Valencia
Icon indicating open access to content
QR code linking to this content
Protocol CitationAlejandro Vignoni, Yadira Boada 2023. PLATERO: A Calibration Protocol for Plate Reader Green Fluorescence Measurements. protocols.io https://dx.doi.org/10.17504/protocols.io.kxygxzjz4v8j/v1
Manuscript citation:
10.3389/fbioe.2023.1104445
License: This is an open access  protocol  distributed under the terms of the  Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: August 02, 2022
Last Modified: January 13, 2023
Protocol  Integer ID: 68080
Keywords: fluorescence calibration, serial dilution, calibration protocol for plate reader green fluorescence measurement, plate reader green fluorescence measurement, fluorescence measure, plate reader fluorescence measurement, fluorescein calibration model, fluorescence intensity value, fluorescence measures in concentration unit, green fluorescent protein, such as the green fluorescent protein, certain fluorescent molecule, calibration protocol, reference fluorophore, measurement device gain, measurement device, synthetic biology, measure of the light, gain effect of the measurement device, variability assessment
Abstract
One of the most common sources of information in Synthetic Biology is the data
coming from plate reader fluorescence measurements. These experiments provide a measure of
the light emitted by certain fluorescent molecules, such as the Green Fluorescent Protein (GFP).
However, these measurements are generally expressed in arbitrary units and are affected by the
measurement device gain. This limits the range of measurements in a single experiment and
hampers the comparison of results among experiments. In this work, we provide a calibration
protocol to express fluorescence measures in concentration units of a reference fluorophore. The
protocol removes the gain effect of the measurement device on the acquired data. In addition,
the fluorescence intensity values are transformed to units of concentration using a Fluorescein
calibration model. Both steps are expressed in a single mathematical expression which returns
normalised, gain independent, and comparable data, even if the acquisition was done at different
device gain levels. The protocol embeds a Lineararity and Bias Analysis that provides an
assessment of the uncertainty of the model estimations, and a Reproducibility and Repeatability
analysis that evaluates the sources of variability originating from the measurements and the
equipment. All the functions used to build the model, exploit it with new data, and perform the
uncertainty and variability assessment are available in an open access repository.
Materials
Fluorescein Sodium Salt 1X PBS (Phosphate-buffered saline )

Protocol materials
Fluorescein Sodium Salt
1X PBS (Phosphate-buffered saline )
Safety warnings
Make sure to record all information about your instrument to document your experiment. If your instrument has variable temperature settings, the instrument temperature should be set to room temperature (approximately 20-25 C) for all measurements.
Before start
Before beginning these protocols, please ensure that you are familiar with the measurement modes and settings of your instrument. For all of these calibration measurements, you must use the same plates and volumes that you will use in your cell-based assays. You must also use the same settings (e.g., filters or excitation and emission wavelengths) that you will use in your cell-based assays. If you do not use the same plates, volumes, and settings, the calibration will not be valid.

Stock Reference Solution
Start from at least1 mL of10 micromolar (µM) Fluorescein Sodium Salt solution in 1X PBS (Phosphate-buffered saline )

[OPTIONAL] If you have access to a spectrophotometer, you can calculate the concentration of your Fluorescein reference stock solution even more accurately using the Beer-Lambert law.

Measure the solution's absorbance at 492 nm and calculate concentration using an extinction coefficient of 68.029 mM-1 cm-1

If the concentration of your stock reference solutions is different than 10 micromolar (µM) then you need to recalculate the dilution of the next step to obtain a Dilution 1 solution of 0.625 micromolar (µM)

Prepare the starting dilution of Fluorescein solution
Dilute 250 µL of the stock reference solution into 3750 µL of 1X PBS (Phosphate-buffered saline ) to obtain 4 mL of Dilution 1 @ 0.625 micromolar (µM)

Prepare the serial dilutions of Fluorescein
Accurate pipetting is essential. Serial dilutions will be performed in 5 tubes. There will be a sixth tube that must contain PBS buffer only. Initially, you will set up 5 tubes labeled Dilution 2 to 5 and PBS, with 2 mL of 1X PBS (Phosphate-buffered saline ) .
Then you will perform a serial dilution by consecutively transferring 2 mL from Dilution 1 to Dilution 2 and so on, always with good mixing (pipetting up and down three times or more).


Label four tubes with the names Dilution 2 to 5, and one tube with the name PBS.
Add 2 mL of 1X PBS (Phosphate-buffered saline ) to all tubes.

Transfer 2 mL of Dilution 1 into Dilution 2 tube. Mix thoroughly by pipetting up and down three times or more.

Transfer 2 mL of Dilution 2 into Dilution 3 tube. Mix thoroughly by pipetting up and down three times or more.

Transfer 2 mL of Dilution 3 into Dilution 4 tube. Mix thoroughly by pipetting up and down three times or more.

Transfer 2 mL of Dilution 4 into Dilution 5 tube. Mix thoroughly by pipetting up and down three times or more.

Transfer 2 mL of Dilution 5 into the liquid waste.

Micro-plate set up
Transfer 100 µL of each dilution into the corresponding well using the following plate map:


AB
PBSA1, A10, B5, B8, C1, C3, D10, D7, E2, E5, F12, F9, G4, G6, G7, H11
Dilution 1A4, A7, B11, B3, C5, D1, D12, D9, E4, E6, F1, F10, F3, H12, H2, H8
Dilution 2A12, A2, B1, B6, C8, D11, D4, E12, E8, F5, F7, G2, G5, G9, H10, H6
Dilution 3A3, A5, B9, C10, C12, C2, C6, D8, E1, E11, F11, F4, F6, G10, G3, H7
Dilution 4A11, A8, A9, B10, B2, B7, C4, D2, D6, E10, E3, G1, G12, G8, H3, H5
Dilution 5A6, B12, B4, C11, C7, C9, D3, D5, E7, E9, F2, F8, G11, H1, H4, H9

Note
You can use any random distribution of well you like, this one is just the distribution we used as an example.

Measure Fluorescence
Measure the fluorescence of all samples in your plate reader, repeat measurements using 4 different gain settings, and then repeat everything 8 times.

Record data
Save your measurements into an excel file with each repetition on a different sheet using the template.
Download Platero_Data_Template.xlsxPlatero_Data_Template.xlsx