Oct 29, 2020

Public workspaceIn-house automated Smart-Seq2

  • 1Wellcome Sanger Institute
  • Human Cell Atlas Method Development Community
Icon indicating open access to content
QR code linking to this content
Protocol CitationLira Mamanova 2020. In-house automated Smart-Seq2. protocols.io https://dx.doi.org/10.17504/protocols.io.bnzsmf6e
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
We use this protocol and it's working
Created: October 27, 2020
Last Modified: October 29, 2020
Protocol Integer ID: 43794
Keywords: RNA-Seq, Automation, SmartSeq2, Zephyr, Mantis
Abstract
This SOP describes the procedure for the in-house automated Smart-Seq2 workflow with off-the-shelf reagents. This set up utilises the benchtop robots and significantly reduces the experimental cost, hands-on time, as well as increase the reproducibility and workflow efficiency, while delivering a throughput of thousands of cells per day.
Following library construction, samples are pooled in equivolume and quantified, prior to sequencing on the Illumina platform.


Materials
Reagents:
Triton™ X-100  (Sigma, cat. no. t9284)
ERCC RNA Spike-In Mix (Ambion, cat. no. 4456740)
RNase free water (not-depc treated) (Ambion, cat. no. AM9937)
RNaseZap (Ambion, cat. no. AM9780)
CD45 MicroBeads, mouse (Miltenyi Biotec, cat. no. 130-052-301)
dNTP Mix (ThermoFisher UK Ltd, cat. no. 11853933)
Recombinant Ribonuclease Inhibitor (Takara Bio, cat. no. 2313B)
Betaine solution, 5 M (Sigma-Aldrich Co. Ltd, cat. no. B0300-5VL)
SmartScribe RT (Takara Bio, cat. no. 639538)
Maxima H Minus Reverse Transcriptase (Thermo Scientific™, EPO752)
MgCl 2 (1 M) (Life Technologies Ltd, cat. no. AM9530G)
KAPA Hifi HS  Mix (Kapa Biosystems, cat. no. KK2603)
Oligo-dT(30)VN (5′–AAGCAGTGGTATCAACGCAGAGTACTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTVN-3′; IDT, PAGE purification)
IS PCR Primers (5′-AAGCAGTGGTATCAACGCAGAGT-3′; IDT, Standard Desalt)
Template-switching oligo (TSO) (5′-AAGCAGTGGTATCAACGCAGAGTACATrGrG+G-3′; EXIQON, HPLC, RNAse-free)
80% Ethanol Freshly prepared Ethanol 95–97% (vol/vol) AnalaR Normapur analytical reagent (VWR, cat. no. 20823.327) ! CAUTION Ethanol is flammable, and it should be stored carefully and handled with appropriate safety equipment.
DTT (Invitrogen, cat. no. 18064-014) ! CAUTION DTT is toxic when ingested. Avoid inhaling fumes or contact with the skin. Handle it using appropriate safety equipment.
RNAClean XP, 40 mL (Beckman Coulter Genomics, cat. no. A63987) ! CAUTION Agencourt AMPure XP Beads contains Sodium Azide, which is Toxic and should be handled with appropriate care.
Buffer RLT (Qiagen, cat. no. 79216) ! CAUTION Buffer RLT contains guanidine thiocyanate, which is harmful and should be handled with appropriate safety equipment.
Agencourt AMPure XP Beads (Beckman Coulter, cat. no. A63882) ! CAUTION Agencourt AMPure XP Beads contains Sodium Azide, which is Toxic and should be handled with appropriate care.
Bleach (VWR International Ltd, cat. no. 1053393)! CAUTION Bleach can cause an irritation in the eyes, mouth, lungs and on skin.
Agilent High-Sensitivity DNA Kit (Agilent Technologies, cat. no. 5067-4626)
Agilent 1000 DNA Kit (Agilent Technologies, cat. no. 5067-1505)
Nextera XT DNA Sample Preparation Kit, 96 samples (Illumina, cat. no. FC-131-1096)
Nextera XT 96-Index Kit, 384 samples (Illumina, cat. no. FC-131-1002)

Equipment:
Seal - Adhesive PCR Film Kbioscience, cat. no. 4TI-0500)
50 Framestar 96-well PCR Plate (4titude, cat. no. 4TI-0960)
Eppendorf twin.tec 96-well PCR Plate (Eppendorf, cat. no. 0030128648)
Thermo Scientific Nunc 96-well deepwell storage plates (Thermofisher, cat. no 278743)
2ml 96 well V-Bot Assay (Corning, cat. no. 3961)
Dna Lobind Tubes, 1,5ml, PCR Clean, (Eppendorf, cat. no. 0030 108.051)
Falcon® Conical Tubes, 50 mL (Stem Cell Technologies SARL, cat. no. 38010)
Falcon® Conical Tubes, 15 mL (Stem Cell Technologies SARL, cat. no. 38009)
Eppendorf Combitips advanced® 25 mL. Pack of 100. Red (Eppendorf, cat. no.  0030089472)
Tips LTS 20UL Filter RT-RT-L10FLR (Anachem, cat. no. 17007957)
Tips GP-LTS-A-10µL-/F-960/10 (Anachem, cat. no. 30389274)
Pipette Tips LQR LTS 20µL FL 960/10 (Anachem, cat. no.17014400)
Tips 5-200ul Filter Greenpak LTS Pk960Tip (Anachem, cat. no. 30389276) 
Tips 50-1000ul filter Greenpak LTS Pk/768 (Anachem, cat. no. 30389272)
80 uL - Barrier Sterile 96 Rack Tips (Perkin Elmer, cat. no. 111624)
150 uL - Barrier Sterile 96 Rack Tips - (Perkin Elmer, cat. no. 111426)
Pipet-Lite Multi Pipette L8-20XLS+(2-20 μL) (Anachem, cat. no. 17014392)
Pipet-Lite Multi Pipette L8-200XLS+(20-200 μL) (Anachem, cat. no. 17014391)
Pipet-Lite LTS Pipette L-1000XLS+ (Anachem, cat. no. 17014382)
Pipet-Lite Multi Pipette L8-10XLS+(0.5-10 μL) (Anachem, cat. no. 17014388)
Multi-dispenser (Multipette Eppendorf, E3f), Multipette Xstream®, with charging adapter, 100-240V/50-60Hz Eppendorf UK Ltd, cat. no. 4986000025)
Zephyr® G3 NGS Workstation (Perkin Elmer, cat. no.133750)
Plate-Magnetic, 96 ring, bead separator (Perkin Elmer, cat. no.128316)
Spacer, Magnetic plate, Sciclone NGS (Perkin Elmer, cat. no.133514)
96S Super Magnet Plate (Alpaqua, cat. no. A001322)
MANTIS® Liquid Handler (Formulatrix, cat. no. MANTV3.1)
Low Volume Silicone based Mantis Chip (Formulatrix, cat. no. MCLS12)
High Volume Silicone based Mantis Chip (Formulatrix, cat. no. MCHS12)
Alpha Cycler 4 (LabTech, cat. no.AC496)
VWR® PCR Workstation (VWR, cat. no. 732-2542)
2100 Bioanalyzer (Agilent Technologies LDA UK Limited, cat. no.G2938C)
DynaMag Spin Magnet (Life Technologies, cat. no. 12320D)
Vortex mixer (e.g., Grant Instruments, model no. PV-1)
Microcentrifuge (e.g., Starlab, model no. N2631-0007)
Fluorescence-activated cell sorter compatible with single-cell deposition into 96-well plates (e.g., BD Influx, Sorter CL2-BD INFLUX or Beckman Coulter MoFlo)
Refrigerated centrifuge and adaptors for 96-well plates (e.g., Eppendorf, model no. 5810 R)
An Illumina next-generation sequencing platform (HiSeq platform)



Preparation of lysis buffer plates and single-cell samples
Preparation of lysis buffer plates and single-cell samples

Note
A single-cell contains a very low amount of RNA. Therefore, extra effort should be taken to avoid sample degradation and reagent contamination. All work surfaces and equipment should be decontaminated with DNA and RNAZap before and after the experiment. RNA related work should be carried out in a specially designated cleanroom or in a laminar flow hood equipped with UV light for sterilization. When working with a few batches of samples, we would advise pre-aliquotting reagents, especially if the reagent supplied in a large volume.

For reagent dispensing steps, we use the Formulatrix Mantis microfluidic liquid handler, and for all purification steps, we use a Zephyr automated liquid-handling robot.

Steps 1-12 should be performed in specially designated RNAse free areas, keeping reagents and cells chilled at all times.

The Formulatrix Mantis microfluidic liquid handler is positioned in a laminar flow hood. The lysis buffer dispensing step can be performed on alternative liquid handling platforms or manually. The reagents were dispensed from a 0.2 ml or 1 ml tip inserted into the High Volume (HV) Silicone based Mantis chip. We recommend having an individual set of chips for various reagent dispensing. Clean the dispensing chips prior to use by dispensing 50-100 μl of solutions into the waste plate in the following order: Water => 10% Bleach (incubate for Duration00:05:00 ) => Water=> 80% EtOH=>Water. Repeat these procedures after finishing the reagent dispensing.


For each well, prepare the Triton buffer by adding Amount0.2 µL of RNAse inhibitor and Amount0.02 µL of diluted (1/500,000) ERCCs to the Amount1.8 µL of Humidity0.2 % Triton X-100. Prepare an RLT lysis buffer by using Amount9.8 µL of RLT buffer (Qiagen); Amount0.2 µL of DTT (2M) and Amount0.02 µL of diluted (1/500,000) ERCCs. Dispense Amount2 µL of Triton lysis buffer or Amount10 µL of RLT lysis buffer into each well of a 96-well plate.

Note
We recommend to seal dispensed plates and centrifuge them immediately (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) to spin down the lysis buffer to the bottom of the well to prevent a sample degradation.

Sealed Lysis buffer plates can be stored at Temperature-80 °C prior to cell sorting for < 6 months.




FACS sort single cells into the plate. Immediately centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) plates containing cells and keep them chilled on ice or at Temperature-80 °C until further processing.

Note
When FACS sorting, take care of plate calibration/priming prior to single-cell deposition. If many plates are deposited at the same time, repeat the calibration/priming at least every 8 plates.We recommend defrost lysis buffer plates on ice prior to cell sorting, centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) and keep chilled on ice.

Samples can be stored at Temperature-80 °C for at least 6months.


1m
Preparation of RNA Elution buffer
Preparation of RNA Elution buffer
Prepare the RNA elution buffer fresh for each experiment, as described below. We would recommend noting the lot numbers for reagents and plastics for every run, as this can be useful in troubleshooting. Keep the RNA elution buffer chilled until use.

ReagentPer well (μl)Final concentration
100 μM oligo dT 0.152.5 μM
10 mM dNTP1.52.5 mM
RNase Inhibitor (40U/μl)0.1510 U/μl
Nuclease-free water4.2
Total6

Depending on the lysis buffer used, continue with option RLT buffer treated cells (A) or Triton buffer treated cells (B)
We recommend introducing the non-template and multicell (10-50 cells) controls and locate them in the middle of the plate.
Note
RLT buffer should be removed prior to subsequent reverse transcription.

A. RLT buffer treated cells: Single-cell RNA cleaning with RNAClean XP beads

i. Allow RNAClean XP beads to warm up to room temperature for Duration00:15:00 before use, gently mixing to ensure that the beads are evenly resuspended. 

ii. Dispense Amount25 µL of RNAClean XP beads into a clean 96-well Bead Plate with Eppendorf Multi-dispenser.

iii. Seal the plate and centrifuge it (Centrifigation1000 rpm for Duration00:01:00 atTemperature4 °C ) to remove air bubbles in the wells.
Note
Keep bead plate covered with a plate seal on a shaker between cleaning procedures if there are many plates to be processed.

iv. Thaw and centrifuge (Centrifigation1000 rpm for Duration00:01:00 atTemperature4 °C ) plate containing lysed cells from Step 2. Add additional RNAse-free water to the samples if required to a final volume of Amount10 µL

Note
We perform all purification steps using a Zephyr automated liquid-handling robot. Expose Zephyr deck to UV light for at least Duration00:20:00 . For the csRNA cleaning procedures use Zephyr 150 μl Barrier Sterile tips.


v . Add Amount22 µL of RNAClean XP beads to each well of the working plate from step iv at room temperature, and mix thoroughly by pipetting up and down on the Zephyr instrument. Allow the mixture to stand for Duration00:05:00 at room temperature.

vi. Move the plate to a low-elution magnet, and allow the beads to settle for Duration00:02:00 or until the solution is clear.

vii. Once the RNAClean XP beads have settled, carefully remove and discard the supernatant without disturbing the beads. Keeping the 96-well plate on the magnet, wash the RNAClean XP beads with Amount100 µL of freshly prepared Humidity80 % (vol/vol) ethanol for Duration00:00:30 , then remove and discard the ethanol wash. Do not disturb the beads.

viii. Repeat Step vii once more.

ix.Dispense Amount6 µL of RNA elution buffer from step 3 to each well of a new clean 96-well plate and place it on a chilled position of the robot deck. Add Amount5 µL of RNA elution buffer to the RNAClean XP beads with purified single-cell RNA using the Zephyr.

CRITICAL STEP Take the plate that contains beads with RNA and elution buffer under the hood and manually resuspend the beads in the RNA elution buffer with low-bind 10 μl tips with a multichannel pipette to facilitate the resuspension. Make sure that all beads are removed from the tips.

x. Seal the plate and centrifuge it (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ).

xi. Proceed directly to the denaturation step 5.

B.Triton Buffer treated cells

i. Thaw and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) plate containing lysed single cells from Step2.

ii. Dispense Amount5 µL of freshly prepared RNA elution buffer from step 3 to each well of 96-well plate from step i. Seal the plate and centrifuge it (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ).

Proceed directly to the denaturation step 5.








27m 30s
RNA denaturation
RNA denaturation
Perform RNA denaturation on the thermal cycler with the heated lid set to 100 °C.
TemperatureTimeCycles
72 ⁰C3 min1
4 ⁰Cforever
When the temperature reaches Temperature4 °C , immediately transfer plate with denatured single-cell RNA onto ice and proceed to the reverse transcription immediately.

Reverse Transcription
Reverse Transcription
Prepare the Reverse Transcription (RT) Master mix, as described below. We used this protocol for two reverse transcriptases, SmartScribe (SMT) and Maxima -H (Maxima), with enzymes corresponding to Master Mix compositions.

RT Master mix for Maxima enzyme
Reagent Per well (μl)Final concentration
Maxima buffer (5x)22.5x
Betaine (5 M)21.5 M
MgCl2 (100 mM)0.913 mM
TSO (10 μM)1.01.5 μM
RNase inhibitor (40 U/μl)0.2510 U/μl
Maxima RNaseH-minus RT (200 U/μl)0.120 U/μl
H2O0.75
Total7.0

RT Master mix for SMT enzyme
Reagent Per well (μl)Final concentration
Smartscribe buffer(5x)22.5x
DTT (100mM)0.58.5 mM
Betaine (5M)21.5 M
MgCl2 (1M)0.090.013 M
TSO (100 μM)0.11.8  μM
RNase inhibitor (40U/ μl)0.251.76 U/μl
SmartScribe (100U/ μl)0.510 U/μl
H2O0.56
Total5.7


Dispense Amount7 µL of Maxima RT master mix or Amount5.7 µL of SMT RT master mix to the plate with denatured RNA from step 6. Seal the plate, vortex on a medium speed and centrifuge it (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ).

1m
Reverse transcription will convert polyA (+) mRNA to cDNA. Place the plate in the thermal cycler and perform the following steps on a thermal cycler:

Reverse Transcription conditions for Maxima RT
TemperatureTimeCycles
50 ⁰C90 min1
85 ⁰C5 min1
4 ⁰CHold1

Reverse Transcription conditions for SMT RT
TemperatureTimeCycles
42 ⁰C90 min1
50 ⁰C2 min10
42 ⁰C2 min
70 ⁰C15 min1
4 ⁰Chold1
Collect the plate from the thermal cycler, centrifuge it (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) and keep chilled on ice. 
1m
PCR Amplification
PCR Amplification
Prepare the PCR amplification Master mix, as described below.
Note
Identify the number of PCR cycles for each investigated cell type to avoid under- or over-amplification. 

PCR amplification master mix for Maxima enzyme
Reagent Per well (μl)Final concentration
ISPCR primer (10 μM)0.50.4 μM
KAPA HiFi Hot Start Ready Mix (2x)12.52x
H2O1.0
Total14

PCR amplification master mix for SMT enzyme
Reagent Per well (μl)Final concentration
ISPCR primer (10 μM)0.250.3 μM
KAPA HiFi Hot Start Ready Mix (2x)12.52x
H2O2.25
Total15
Dispense Amount14 µL of PCR master mix for Maxima RT or Amount15 µL of PCR master mix for SMT RT to each sample on the 96-well plate from step 10 using a Mantis. Seal the plate, vortex on a medium speed and centrifuge it (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ). Place the plate in the thermal cycler and perform the following steps:

PCR amplification protocol for Max RT
TemperatureTimeCycles
98 ⁰C3 min1
98 ⁰C15 sec20 / 22 / 25 cycles depending on the cell type
67 ⁰C20 sec
72 ⁰C6 min
72 ⁰C6 min1
4 ⁰Chold

PCR amplification protocol for SMT RT
TemperatureTimeCycles
98 ⁰C3 min1
98 ⁰C20 sec20 / 22 / 25 cycles depending on the cell type
67 ⁰C15 sec
72 ⁰C6 min
72 ⁰C5 min1
4 ⁰Chold

Note
Amplified cDNA samples can be stored in Temperature-20 °C for several months before purification.



1m
Proceed with the Agencourt AMPure beads for cleaning on the Zephyr instrument, or keep the plate at Temperature4 °C if continuing the next day, or Temperature-20 °C for long term storage.

Purification of the amplified cDNA
Purification of the amplified cDNA
15m
15m

Note
Steps 14–25should be performed in a dedicated post-amplification room.
Allow Agencourt AMPure XP beads to warm up to room temperature for Duration00:15:00 before use, gently mixing to ensure that the beads are evenly resuspended. Dispense Amount25 µL of Agencourt AMPure XP beads into the clean 96-well Bead Plate with Eppendorf Multi-dispenser.

15m
Centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) the plate containing the amplified cDNA from Step13.
1m
Add Amount25 µL of the Agencourt AMPure XP beads to each well of the 96-well plate containing the PCR-amplified cDNA molecules at room temperature, and mix thoroughly by pipetting up and down. Allow the mixture to stand for Duration00:05:00 at room temperature.

5m
Move the plate to a magnet, and allow the beads to settle for Duration00:02:00 or until the solution is clear.

2m
Carefully remove and discard the supernatant without disturbing the beads, once the beads have settled.
Keeping the 96-well plate on the magnet, wash the beads with Amount100 µL of freshly prepared Humidity80 % (vol/vol) ethanol for Duration00:00:30 , and then remove and discard the ethanol wash. Do not disturb the beads.

30s
Repeat Step 19 once more.
Remove and discard any remaining ethanol solution from the well without disturbing the beads, and then allow the beads to dry for ~Duration00:05:00 .

5m
Add Amount25 µL of nuclease-free water to the beads, remove the plate from the magnet and resuspend by pipetting up and down at room temperature.

Incubate this 96-well plate for Duration00:02:00 off the magnet at room temperature.

2m
Return the 96-well plate to the magnet, and allow the beads to settle for Duration00:05:00 or until the solution is clear.

5m
Transfer the supernatant, which contains the purified cDNA to a new 96-well plate at room temperature without disturbing the beads.

Note
The amplified cDNA can be stored at Temperature-20 °C for >6 months before library preparation.


Quality Control of Amplified cDNA and input normalization
Quality Control of Amplified cDNA and input normalization
We use an Agilent Bioanalyzer for a cDNA QC on a high-sensitivity chip. Run Amount1 µL of the eluted sample on a high-sensitivity chip on the Agilent. Dilute the library to 1 in 10 if cDNA is highly concentrated. It is important to quantify cDNA concentration accurately. Overestimation can affect the normalization of the cDNA input for the Nextera library preparation and lead to skewed Nextera libraries profiles.

Sample normalization
Sample normalization

Note
Steps 27–29 should be performed in a dedicated post-amplification room. Illumina libraries are prepared using the tagmentation approach with Nextera XT DNA Library Prep Kit (Illumina). We recommend following the manufacturer’s instructions for the applied kit.
For cDNA input normalization, we check 11 randomized wells throughout a plate (A6; B4; B11; C3; D4; D12; F2; F9; G3; G8; H5). Measure cDNA concentration between 200-8000bp in all 11 samples, and find an average concentration that will be applied to the whole plate. Remove the outliers for accurate template normalization. The final average concentration of cDNA should be normalized to 0.3 ng/μl for optimal tagmentation. Normalize multicell controls separately.
Sample normalization is performed on the Zephyr instrument. A fixed amount of cDNA will be used for the dilution. Take Amount5 µL of cDNA from the individual wells and dilute with water to achieve a final concentration of 0.3 ng/ul.

Seal the normalised plate and centrifuge it (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ), vortex on medium speed for Duration00:01:00 and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) again.

Note
Normalised cDNAcan be stored at Temperature+4 °C for continuation the next day, or at Temperature-20 °C for a week. 


3m
Library construction using the Nextera XT Library Preparation Kit
Library construction using the Nextera XT Library Preparation Kit
Transfer Amount1.2 µL of normalised samples from step 29 using the Zephyr instrument to the working Nextera Plate. Keep plates chilled and continue with the Nextera protocol.

Prepare the Nextera master mix, as described below and dispense with the Mantis using a Low Volume (LV) chip.
Reagent Per well (μl)
Tagmentation DNA buffer2.5
Amplification Tagment Mix1.25
Total3.75
Seal the plate and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ), vortex on medium speedfor Duration00:01:00 and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) again.

3m
Place the Nextera plate in a thermal cycler and run the following program:
TemperatureTime
55 ⁰C10min
4 ⁰CHold

Once the temperature reaches Temperature4 °C , remove from the thermal cycler and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) the plate.

1m
Wash Mantis LV chip with water by dispensing 50-100 μl of water into the waste plate.
Dispense Amount1.2 µL NT buffer into each well to neutralize the reaction.

Nextera Index preparation
Nextera Index preparation
1m
1m

Note
Prepare the Indexes set mix (A-B-C-D) in 96-well Index Plate according to the manufacturer’s instructions. Defrost and spin down Index 1(i7) and Index 2 (i5). Arrange the index primers in the Index Plate Fixture as follows: Index 1 (i7) adapters in columns 1-12 in ascending or descending order. Index 2 (i7) adapters in columns A-H in ascending or descending order(Figure S3).  
Using a multichannel pipette, add Amount10 µL of i7 adapters down each row. Replace the vial cap with a new cap. Using a multichannel pipette, add Amount10 µL of i5 adapters down each column. Replace the vial cap with the new cap.


Seal the plate and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ), vortex on medium speedfor 1min and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) again.
2m
Keep premade plates with index barcodes chilled until use or in the freezer at Temperature-20 °C for long term storage.

Dispense Amount2.5 µL of Index Adapters on the Zephyr instrument to the working Nextera plate from step 36. Seal the plate and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ).

1m
Nextera Library amplification
Nextera Library amplification
40.Wash Mantis LV chip by dispensing 50-100 μl of water into the waste plate.
Dispense Amount3.7 µL of NPM buffer by the Mantis dispenser to each well of the working Nextera plate.

Seal the plate and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ), vortex on medium speed for 1min and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) again.
2m
Place the library prep plate in a thermal cycler and perform the following program:
TemperatureTimeCycles
72 ⁰C3 min1
95 ⁰C30 sec1
95 ⁰C10 sec 12
55 ⁰C30 sec
72 ⁰C30 sec
72 ⁰C  5 min1
4 ⁰CHold1

Once the temperature reaches Temperature4 °C , remove from the thermal cycler. Centrifuge the plate (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ), vortex on medium speed for 1min and centrifuge (Centrifigation1000 rpm for Duration00:01:00 at Temperature4 °C ) again.

2m
Nextera library pooling and purification
Nextera library pooling and purification
Collect Amount5 µL of each sample (except the Control wells) into 8-well PCR strip tubes using a multichannel pipette. Pool control wells if these samples are used for the analysis.

Then pool amplified libraries from PCR strips into a single 1.5 ml vial (≈Amount595 µL ). Briefly spin down the strips and collect the rest of the samples.

Measure the sample volume with the pipette and add Agencourt AMPure XP beads at 1:1 ratio. Mix well by pipetting. Incubate for Duration00:10:00 at room temperature.

10m
Place the vial with the beads/sample mix on the magnet for Duration00:05:00 or wait till the solution is clear.

5m
Carefully remove and discard the supernatant without disturbing the beads, once the beads have settled.
Keeping the vial on the magnet, wash the beads with Amount100 µL of freshly prepared Humidity80 % (vol/vol) ethanol for Duration00:00:30 , and then remove and discard the ethanol wash. Do not disturb the beads.

30s
Repeat Step 50 once more.
Remove and discard any remaining ethanol solution from the well without disturbing the beads, and then allow the beads to dry for ~Duration00:05:00 . Leave the vial cap open to air-dry the beads. 

5m
Remove the vial from the magnet, add Amount100 µL of nuclease-free water to beads and resuspend by pipetting up and down at room temperature.

Incubate the vial for Duration00:02:00 off the magnet at room temperature.

2m
Return the vial to the magnet, and allow the beads to settle for Duration00:05:00 or until the solution is clear.

5m
Transfer the supernatant, which contains the purified Nextera libraries to a new vial without disturbing the beads.
Keep the samples at Temperature4 °C if continued the next day or Temperature-20 °C for long term storage.

Quality Control of Nextera sequencing libraries and normalization for sequencing
Quality Control of Nextera sequencing libraries and normalization for sequencing
Run Amount1 µL of the eluted sample from step 57 in triplicate on an Agilent Bioanalyser using a DNA 1000 assay chip or 1 in 10 diluted sample on a high-sensitivity chip. 

Note the molarity of each reading and take an average of the 3 triplicates discarding any outliers.
Taking the molarity reading from step 59. In a fresh 1.5 ml tube combine the calculated volume of the amplified library pool with nuclease-free water to produce a final sequencing library pool of 2.8 nM.
In a fresh 1.5 ml tube equimolar pool the 2.8 nM sequencing library from step 60 with 3 additional sequencing library pools which have been processed through this workflow. To produce a final sequencing pool containing 384 samples.
Library sequencing
Library sequencing

Libraries were submitted for sequencing on an Illumina HiSeq 4000 chemistry (paired-end, 75-bp reads). Perform single-end (SE) or paired-end sequencing (PE) of the Single-cell derived libraries according to the manufacturer’s protocol. Alternative sequencing options to achieve comparable sequencing depth are; the NextSeq 550 with high output flow cell or NovaSeq 6000 SP flow cell on one lane of an XP workflow.