1. If the machine is running, check the “tuning log sheet.ods” and get the date of the last tuning. If
the last tuning was done more than two days ago proceed to step 3.3.
2. Check argon supply and pressure (5 bar). Make sure that one of the argon tanks has at least
25% which will last overnight. Get a new tank if required from the shop in the basement using the
3. Check Helium pressure (2 bar). Check the Helium bottles in the cabinet. If the pressure drops
below 100 bar open the second bottle and close the one that was open. Move the empty bottle in
front of the room and hang a sign “Empty” on it. Inform Esther Danenberg or Stefanie Engler to
4. Open the CyTOF software on the computer and log in with your name and password.
5. Insert the tuning slide into the Hyperion.
7. Wait for at least 30 minutes until you proceed with tuning.