May 08, 2019

Public workspaceiGEM Calibration Protocol - Flow Cytometry Fluorescence

  • 1iGEM Measurement Committee
  • iGEM Measurement
    Tech. support email: pauljrutten@gmail.com
Icon indicating open access to content
QR code linking to this content
Protocol CitationJacob Beal, Cheryl Telmer, Richard Tennant, Paul Rutten 2019. iGEM Calibration Protocol - Flow Cytometry Fluorescence. protocols.io https://dx.doi.org/10.17504/protocols.io.2pcgdiw
License: This is an open access protocol distributed under the terms of the Creative Commons Attribution License,  which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Protocol status: Working
Created: May 07, 2019
Last Modified: August 07, 2019
Protocol Integer ID: 22980
Keywords: iGEM, Flow cytometry, Fluorescence
Abstract
This protocol can be applied to any strain of cell that can be safely run through a flow cytometer. For clarity, we have written it assuming E. coli DH5-alpha; to apply the protocol to another cell type, substitute the other cell type for any place where the protocol says [E. coli DH5-alpha].

This protocol has been written for measurement of GFP, YFP, or other yellow/green fluorescent proteins into MEFL units. To apply it to fluorescent proteins of other colors:

  • Replace BBa_J364001 with a construct for strong expression of the other protein.

  • For blue proteins (e.g., mTagBFP), measure with 405nm excitation and 450nm/50nm emission filter. Units will be MEC30.

  • For red/orange proteins (e.g., mCherry), measure with 561nm excitation and 610nm/20nm or 620nm/15nm emission filter.  Units will be MEPTR.

  • For far-red / near-infrared proteins (e.g., IRFP), measure with 635nm excitation and 780nm/60nm or 750nm long-pass (LP) emission filter. Units will be MEAPCY7.

To apply the protocol to multiple colors, add a positive process control for each color and use one of the tools on the iGEM Measurement Resources page to compensate measurements for spectral overlap.

This protocol can be combined with bead-based cell size calibration.
Materials
MATERIALS
ReagentRainbow calibration beadsSpheroTechCatalog #RCP-30-5A
ReagentRainbow calibration beadsSpheroTechCatalog #URCP-38-2K
ReagentRainbow calibration beadsSpheroTechCatalog #URQP-38-6K
  • Only one set of rainbow calibration beads are necssary, options are listed.

  • Negative process control: wild-type [E coli DH5-alpha]

  • Positive process control: [E coli DH5-alpha] transformed with BBa_J364001 (or another similar construct for strong constitutive GFP expression)

  • Experimental samples: [E coli DH5-alpha] transformed per your desired experimental procedure

  • Flow cytometer with a channel that measures with a 488nm excitation and a 530nm/30nm emission filter. Measurements on this channel will be converted to MEFL.


Download iGEM 2019 Flow Cytometry Fluorescence Calibration.xlsxiGEM 2019 Flow Cytometry Fluorescence Calibration.xlsx

Download iGEM 2019 Flow Cytometry Fluorescence Calibration - Example.xlsxiGEM 2019 Flow Cytometry Fluorescence Calibration - Example.xlsx


Before start
Read through this entire protocol carefully before you start your experiment and prepare any materials you may need. See the "Results" section for an example of a completed data analysis spreadsheet.
Acquisition of Data
Acquisition of Data
Prepare and culture experimental samples, negative process control, and positive process control according to your desired experimental procedure.
Prepare experimental samples, negative process control, and positive process control as needed for running through your flow cytometer.
Shake SpheroTech rainbow calibration beads vigorously or vortex briefly. Prepare bead sample by diluting 15 μL of SpheroTech rainbow calibration beads into 200 μL of 1x PBS.
Measure all samples in flow cytometer

  • Using the bead sample, adjust MEFL channel voltage so that the top bead peak is below detector saturation.

  • Using the negative process control sample, adjust forward-scatter and side-scatter voltages to place the strong cell peak as close to the center of the detector range as possible (see example illustrated below).



  • Instrument gating should be set to ensure that no cell events are discarded (see example illustrated below).



  • Collect at least 10,000 events per sample.
Compute the MEFL/a.u. scaling factor from bead sample either using one of the tools on the Measurement resources page or following the Excel-based scaling factor computation below.
Multiply all fluorescence measurements by MEFL/a.u. scaling factor to convert from a.u. to MEFL.
Excel-based scaling factor computation
Excel-based scaling factor computation
Compute a histogram of GFP channel measurements for the rainbow calibration bead sample; this should show some number of distinct high peaks (as in the example below).



Identify the center of each distinct, narrow, high peak, and record these locations in the provided cell of the Fluorescence Calibration spreadsheet, starting with the highest peak. Leave all other locations blank. Download iGEM Data Analysis Template - Flow Cytometry Fluorescence Calibration - v1.xlsxiGEM Data Analysis Template - Flow Cytometry Fluorescence Calibration - v1.xlsx
Select the model and channel for your calibration beads from the menu in the sheet.
Record the lot number for your calibration beads (e.g., “AJ02”) in the provided cell. This should cause a set of bead values to appear. The lot number must precisely match the lot in the “Bead Catalog” sheet.
The MEFL/a.u. scaling factor and the graph on the sheet should now fill in.
The graph shows one point for each peak and a line for the MEFL/a.u. scaling factor.
Check that the points are all close to the line (as in the example below). If this is not the case, check to be sure that you have identified and recorded your peaks correctly.