a. Cleaning the Tubing and the Pathway
i. Make sure the stream is on.
ii. Load and Run ~ 10 ml of 20% bleach for 10-15 minutes, with the flow rate set to 10.
iii. Remove the tube of bleach.
iv. Load and Run ~ 10 ml of DI water for 10-15 minutes, with the flow rate set to 10.
v. Unload the DI H2O, and turn off the stream.
b. Cleaning the Flow Cell
i. Open the hood of the cytometer. Turn the nozzle-locking lever counter-clockwise to the 6:00 position and remove the nozzle.
ii. Remove the current nozzle tip and replace it with the metal nozzle (Closed Loop Nozzle) that has the waste line attached to it. Be sure that the nozzle has an O-ring and is oriented with the word TOP facing up.
iii. Turn the black key to the right to lock the nozzle in place.
iv. Select Cytometer > Cleaning Modes > Clean Flow Cell.
v. When prompted, install a tube containing approximately 3 mL of sterile, filtered DI water, then click OK. The cytometer loads the tube and fills the flow cell with the DI water.
vi. Click OK when the completion dialog opens.
vii. Repeat step-iv to step-vi at least two more times.
c. Perform Fluidic Shutdown
i. Make sure the stream is off. Unload the sample tube, if one is loaded.
ii. Check the ethanol shutdown tank and refill if it needed.
iii. Select cytometer> Fluidics Shutdown.
The Fluidics Shutdown dialog opens.
iv. Remove the nozzle from the flow cell assembly and click Done.
v. Insert the integrated closed-loop nozzle into the flow cell assembly and click Done. Verify that there is an O-ring in the nozzle before installing it.
vi. Connect the fluid and the air tube to the stainless steel ethanol (ETOH) shutdown tank.
1. Disconnect the fluid line from the bottom side of the sheath filter and connect it to the ethanol filter on the ETOH shutdown tank.
a. Keep the sheath filter attached to the sheath tank. Do not run ethanol through the sheath filter
2. Disconnect the air line from the sheath tank and connect it to the air port on the ETOH shutdown tank.
3. Click Done. The system starts the cleaning process and then displays a message at the bottom of the dialog
vii. When prompted, install a tube containing 3 mL of sterile, filtered DI water on the loading port, then click Done.
1. The cytometer loads the tube and continues the cleaning process. A progress message appears, and then displays Done when the process is complete.
viii. Click OK when you see a message informing you that the system can be turned off.
ix. Vent the air pressure from the sheath tank by pulling up on the ring on the pressure relief valve
x. Turn off the cytometer main power.
xi. Exit BD FACSDiva software and shut down the computer.