Add smart lysis buffer with 1:1 ratio in the VTM sample


Heat for 68C for 8 minutes and then 95C for 10 minutes
Cool and take 3 ul to colorimetric LAMP master mix (NB: M1800)
Vortex reaction mix and centrifuge to collect material
Pipet 24 μl per reaction into desired reaction vessels and add 1 µl of sample. Mix by vortexing or by pipetting if using a plate or similar vessel, centrifuge to collect if necessary. Check that reaction solutions have a bright pink colour, which indicates initial high pH required for successful pH-LAMP reaction.

Incubate at 65°C for 30 minutes.

Cool on ice for 2 minutes
Add 5 ul of the LAMP reaction with the sample buffer
Load to the LFD strip and let is absorbe
Visualize the bands for internal control and SARS-CoV-2
Scan with mReader to record results